Limit Function - Honeywell APT4000PH User Manual

Apt4000 series, 4-wire ph analyzers
Table of Contents

Advertisement

Configuration

Limit function

Relay 1
Menu group
Code Display
Relay / Controller
rL.
52
Select menu item
L1.
Contact function
Contact response
Enter switching point
Enter hysteresis
Delay
L2.
Relay 2 menu group
Ct.
Controller menu group
End:
Press CONF, then ENTER
APT4000PH
Code Display
Action
Use of relays:
rL.
• Limit function (LiMIT)
• Controller (CtROL)
Select with
proceed with ENTER
Note: Selecting CtROL leads to
Controller menu group Ct.
L1.
Limit 1 function , see Page 55.
Select with
proceed with ENTER
Limit 1 contact response
N/C: normally closed contact
N/O: normally open contact
Select with
proceed with ENTER
Limit 1 switching point
Select with
key, proceed with ENTER
Limit 1 hysteresis
Select with
key, proceed with ENTER
Limit 1 delay
The contact is activated with
delay (deactivated without delay)
Select with
key, proceed with ENTER
Choices
LiMIT
(CtROL)
key,
Lo
(Hi)
key,
N/O
(N/C)
key,
00.00 pH
(
XX.XX pH
key, edit number with
XXXX mV)
00.50 pH
(
key, edit number with
XX.XX pH
XXXX mV)
0010 SEC
(
0 to
9999 SEC)
key, edit number with
53

Hide quick links:

Advertisement

Table of Contents
loading