Configuration
Control of rinsing and calibration probes
Menu group
Code Display
Rinsing and cali-
Pb.
bration probes
Code Display
Action (rinsing probe)
Control of:
Pb.
• Rinsing probe (rinse)
• Calibration probe (EasyClean)
Select with
Proceed with ENTER
Rinsing interval
Select with
with
Rinse duration
Select with
with
Contact type
Select with
Proceed with ENTER
58
Select menu item
Rinsing/calibration probe
Rinsing interval
Rinse duration
Contact response
Cleaning interval
Calibration interval
Selection
rinse
(EASYCLN)
EASYCLN:
key
see opposite
page
000.0 h
key, edit number
(
XXX.X h)
key, proceed with ENTER
0060 s
(0000 to
key, edit number
key, proceed with ENTER
1999 SEC)
N/O
key
(
N/C)
APT4000PH
Action (calibration probe)
Code Display
Calibration probe (EasyClean)
Pb.
Select with
Proceed with ENTER
Cleaning interval
(EasyClean only)
Select with
with
Calibration interval
(EasyClean only)
Select with
with
Lock cleaning (calibration)
interval*
On: The APT4000PH only starts a
cleaning (calibration) interval if the
measured value lies within the toler-
ated range (Limit Lo/Limit Hi).
*"Lock cleaning (calibration) interval" function:
ppb
Locked
Limit Hi
Limit Lo
Locked
Selection
EASYCLN
key
000.0 h
(
XXX.X h)
key, edit number
key, proceed with ENTER
000.0 h
(
XXX.X h)
key, edit number
key, proceed with ENTER
Off
(
On)
The APT4000PH only starts a
cleaning (calibration) interval if the
measured value lies within the
tolerated range (Limit Lo/Limit Hi).
(For limit setting, refer to Pages 53,
55)
t
59