Honeywell APT4000PH User Manual page 21

Apt4000 series, 4-wire ph analyzers
Table of Contents

Advertisement

Configuration
Output 2
Temperature unit and probe. Output current.
Menu group
Code Display
Output 2
o2.
40
Select menu item
Select °C/°F
Select temperature probe
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA for temp error
Hold mode
End:
Press CONF, then ENTER
APT4000PH
Code Display
Action
o2.
Specify temperature unit
Select with
Proceed with ENTER
Select temperature probe
Select with
Proceed with ENTER
Set output current range
Select with
Proceed with ENTER
Current beginning:
Enter lower end of scale.
Select with
key, proceed with ENTER
Current end:
Enter upper end of scale.
Select with
key, proceed with ENTER
Process temperature: Current beginning and current end
Example 1: Range 0 to 100 °C
[°C]
Process temperature
100
Output current
0
[mA]
4
20
Choices
°C
(°F)
key
NTC8.55
(NTC30,
key
PT100,
PT1000)
4–20 mA
key
(0–20 mA)
000.0 °C
(XXX.X)
key, edit number with
100.0 °C
(XXX.X)
key, edit number with
Example 2: Range 50 to 70 °C.
Advantage: Higher resolution in
range of interest
Process temperature
[°C]
70
50
4
20
41
[mA]

Hide quick links: