Honeywell APT4000PH User Manual page 20

Apt4000 series, 4-wire ph analyzers
Table of Contents

Advertisement

Configuration
Output 1
Output current during Error and Hold.
Menu group
Code
Display
Output 1
o1.
38
Select menu item
Select measured variable
Select electrode type
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA in the case of error
Hold mode
End:
Press CONF, then ENTER
APT4000PH
Code Display
Action
22 mA signal for error message
o1.
Select with
Proceed with ENTER
Output signal during Hold
LAST:
During Hold the last measured value
is maintained at the output
FIX:
During Hold a value (to be entered) is
maintained at the output.
Select with
Proceed with ENTER
Only with FIX selected:
Enter current which is to flow at the
output during Hold
Select with
edit number with
proceed with ENTER
Output signal for Hold:
Output current
Output signal Hold
[mA]
Setting FIX = 21.0 mA
21
4
Hold active
Choices
OFF
key
(ON)
LAST
(FIX)
key
021.0 mA
(000.0 to
021.0 mA)
key,
key,
Output signal HOLD
Setting LAST
Hold active
39