Output 2 - Honeywell APT4000PH User Manual

Apt4000 series, 4-wire ph analyzers
Table of Contents

Advertisement

Configuration
Output 1
Output current range. Current beginning. Current end.
Menu group
Code
Display
Output 1
o1.
34
Select menu item
Select measured variable
Select electrode type
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA in the case of error
Hold mode
End:
Press CONF, then ENTER
APT4000PH
Code
Display
Action
o1.
Set output current range
Select with
Proceed with ENTER
Current beginning
Enter lower end of scale, depending
on measured variable selected
(pH or ORP)
Select with
edit number with
proceed with ENTER
Current end
Enter upper end of scale, depending
on measured variable selected
(pH or ORP)
Select with
edit number with
proceed with ENTER
Assignment of measured values:
Current beginning and current end
Example 1: Range pH 0 to 14
[pH]
14
Output current
0
4
20
[mA]
Choices
4–20 mA
(0–20 mA)
key
00.00 pH
(pH:
-2 ... 16)
/ (ORP:
key,
-1500 mV
key,
to
+1500 mV)
14.00 pH
(pH:
-2 ... 16)
/ (ORP:
key,
-1500 mV
key,
to
+1500 mV)
Example 2: Range pH 5 to 7
Advantage: Higher resolution in
range of interest
[pH]
7
Output current
5
[mA]
4
20
35