Honeywell APT4000PH User Manual page 22

Apt4000 series, 4-wire ph analyzers
Table of Contents

Advertisement

Configuration
Output 2
Time constant of output filter.
Menu group
Code Display
Output 2
o2.
42
Select menu item
Select °C/°F
Select temperature probe
Select 0-20 / 4-20 mA
Enter current beginning
Enter current end
Set output filter
22 mA for temp error
Hold mode
End:
Press CONF, then ENTER
APT4000PH
Code Display
Action
o2.
Time constant of output filter
Default setting: 0 s (inactive).
To specify a time constant: Select
with
key, proceed with ENTER
Time constant of output filter
To smoothen the current output 2, a low-pass filter with
adjustable filter time constant can be switched on. When there
is a jump at the input (100 %), the output level is 63 % after
the time constant has been reached.
The time constant can be set from 0 to 120 s.
If the time constant is set to 0 s (default), the current output fol-
lows the input.
Note:
The filter only acts on the current output, not on the display!
Time constant 0 to 120 s
Choices
0 SEC
(0 to
120 SEC)
key, edit number with
43