3 Hydraulic Specifications
3.1.1 Pressure-Flow Curves
0.0
1.0
3.0
2.5
2.0
[bar]
1.5
1.0
0.5
0.0
0.0
2.5
3.2 General Environmental Conditions
Usage
Altitude
Operating ambient temperature
Storage ambient temperature
Operating humidity range
(relative humidity)
Storage humidity range
(relative humidity)
DC supply fluctuations
Transient over-voltages typically present
on the mains supply
Pollution degree
PL-4012-00, Rev01, DCO# 09-188
[gallons/min]
2.0
3.0
10000 rpm
Supply Voltage: 24 VDC
5.0
7.5
10.0
[liters/min]
Figure 4: Pressure/flow rate curves
(Measured with pumphead LPP-200.1)
Table 4: Environmental conditions for pump system
4.0
Specific gravity = 1 g/cm
Viscosity = 0.7 cP
Liquid Temp.: 40C
Speed Limit for 24V ≈ 9500 rpm
Speed Limit for 48V = 10000 rpm
9000 rpm
Supply Voltage: 48 VDC
8000 rpm
5000 rpm
12.5
15.0
17.5
Indoor
(motor with pump head can be placed outdoor)
Up to 2000 m
0
0 to 40
C
0
-20 to 80
C
15 – 95% (non condensing)
15 – 95% (non condensing)
± 10% of nominal voltage
Surge immunity according to EN 61000-4-5
(tested with AC/DC power supply TSL 120-124 for 24VDC
and TIS 600-148 for 48 VDC)
2
User Manual for BPS-200
www.levitronix.com
5.0
6.0
3
7000 rpm
6000 rpm
20.0
22.5
25.0
40
30
[psi]
20
10
0
7
Need help?