4.3 Temperature Monitoring
To avoid overheating of the system, the controller and motor temperatures are monitored. If the controller
temperature exceeds 70°C (158°F) or the motor temperature 90°C (194°F) for a duration of more than 10
minutes, the system goes into an error state and the pump stops. At 80°C (176 F) controller temperature or
100°C (212°F) motor temperature, the system stops immediately.
WARNING
YES
ERROR
YES
ERROR
YES
Figure 9: Controller temp. monitoring
PL-4012-00, Rev01, DCO# 09-188
[gallons/min]
1.0
2.0
10000 rpm
5.0
7.5
10.0
Figure 8: Electrical power consumption
(Controller LPC-200.x with pump head LPP-200.1 and motor BSM-1.x)
Start
Temperature
Monitoring
Temperature
NO
> 70°C ?
YES
Temperature
higher than 70°C
for more than 10
minutes?
NO
Temperature
NO
> 80°C ?
3.0
4.0
9000 rpm
8000 rpm
12.5
15.0
[liters/min]
WARNING
ERROR
ERROR
Figure 10: Motor temperature monitoring
User Manual for BPS-200
www.levitronix.com
5.0
6.0
— Supply = 24 VDC
- - - Supply = 48 VDC
7000 rpm
6000 rpm
5000 rpm
Specific gravity = 1 g/cm
Viscosity = ~ 0.7 cP
Liquid Temp.: 40C
17.5
20.0
22.5
Start
Temperature
Monitoring
Temperature
YES
> 90°C ?
YES
Temperature
higher than 90°C
YES
for more than 10
minutes?
NO
Temperature
YES
> 100°C ?
3
25.0
NO
NO
11
Need help?
Do you have a question about the BPS-200 and is the answer not in the manual?