Advertisement

Operational Characteristics

Substitute Gas Calibration

The infrared gas transmitter Dräger PIR 3000 can also be used to measure other
gases and vapours. The following table shows the required information (see also
"Calibration" on page 11).
1)
Measured gas
acetone
i-butane
n-butane
ethanol
ethyl acetate
ethyl acetate
n-hexane
methanol
n-nonane
n-Octane
n-pentane
i-propyl alcohol
propene (propylene)
toluene
___________
1) The measuring function for the explosion protection according to EN 60079-29-1 is proven, see EC-Type Examination certificate
BVS 05 ATEX E 143X and associated additions.
2) The LEL values were used according to IEC 60079-20-1. Other LEL values may apply for the device settings at the location of use.
3) Typical tolerance: ±5 %.
4) When the following substances are measured at concentrations above 70 % LEL, the deviations of the measured values exceed the
permitted deviations in accordance with EN 60079-29-1.
16
CAS-No.
Measuring range
[%LEL]
67-64-1
0 to 100
75-28-5
0 to 100
106-97-8
0 to 100
64-17-5
0 to 100
141-78-6
0 to 100
141-78-6
0 to 100
110-54-3
0 to 100
67-56-1
0 to 100
111-84-2
0 to 100
111-65-9
0 to 100
109-66-0
0 to 100
67-63-0
0 to 100
115-07-1
0 to 100
108-88-3
0 to 100
1)
Gas Category
Substitute Gas
ethylene
propane
propane
4)
propane
ethylene
4)
propane
propane
4)
ethylene
propane
propane
propane
propane
ethylene
ethylene
Calibration
Response time
2) 3)
factor
t
0...50
0.7
≤ 24 s
1.6
≤ 21 s
1.2
≤ 23 s
0.9
≤ 21 s
0.4
≤ 35 s
1.4
≤ 35 s
1.8
≤ 32 s
0.2
≤ 21 s
1.9
≤ 89 s
1.8
≤ 67 s
1.5
≤ 28 s
1.3
≤ 24 s
0.4
≤ 19 s
0.6
≤ 49 s

Hide quick links: