Explosion-Protection Approvals - Dräger PIR 3000 Instructions For Use Manual

Infrared gas transmitter
Table of Contents

Advertisement

Explosion-Protection Approvals

The explosion-protection approvals are valid for use of the device in gas/vapour-air mixtures of combustible gases and
vapours under atmospheric conditions. The explosion-protection approvals are not valid for use in oxygen-enriched
atmospheres. Unauthorised opening of the enclosure invalidates the explosion-protection approval.
— ATEX
— IECEx
— UL
(Underwriters Laboratories Inc.)
— CSA
1) The year of manufacture is indicated by the third letter of the serial number on the type plate: X = 2006, Y = 2007, Z = 2008,
A = 2009, B = 2010, C = 2011, D = 2012, etc.
example: serial number ARXH-0054, the third letter is U, so the year of manufacture is 2006.
The appendix has a copy of the ATEX certificate, which offers relevant information on safety regulations under subjects such
as "Subject matter and type", "Description", "Characteristic quantity" and "Terms/conditions of safe use".
Type IDS 0001
II 2G Ex db IIC T6 Gb
II 2D Ex tb IIIC T80°C Db IP6X
–40 °C ≤ Ta ≤ +65 °C
Type ITR 001X
II 2G Ex db IIC T6 Gb
II 2D Ex tb IIIC T80°C Db IP6X
–40 °C ≤ Ta ≤ +60 °C
Types IDS 0011 and ITR 000X:
II 2G Ex db eb IIC T6 Gb
II 2D Ex tb IIIC T80°C Db IP6X
–40 °C ≤ Ta ≤ +65 °C
year of construction (by serial number)
BVS 05 ATEX E 143X
Type IDS 0001:
Ex db IIC T6, Ex tb IIIC T80°C Db IP6X
o
o
–40
C ≤ Ta ≤ +65
C
Type ITR 001X
Ex db IIC T6 Gb, Ex tb IIIC T80°C Db IP6X
–40 °C ≤ Ta ≤ +60 °C
Types IDS 0011 and ITR 000X:
Ex db eb IIC T6, Ex tb IIIC T80°C Db IP6X
–40 °C ≤ Ta ≤ +65 °C
BVS 05.0011X
Type IDS 0001:
Class I, Div. 1, Groups A, B, C, D
Class II, Div. 1, Groups E, F, G
Type IDS 0001:
Class I, Div. 1, Groups A, B, C, D
C22.2, No. 152
Explosion-Protection Approvals
0158
0158
0158
1)
C
US
®
LR97594
5