Error Messages - GE Transport PT878 Abridged Manual

Portable liquid ultrasonic flowmeter
Hide thumbs Also See for Transport PT878:
Table of Contents

Advertisement

August 2008

Error Messages

Error
Problem
E0
No error
E1
Low Signal - Poor
ultrasonic signal strength
E2
Sound Speed Error
E3
Velocity Range - Velocity
exceeds programmed
limits
E4
Signal Quality
E5
Amplitude Error
E6
Cycle skip, Acceleration
E7
Analog Out Error
E8
Temperature Input
Supply
E9
Temperature Input Return
3-12
To indicate a possible problem during a measurement, Error
Messages may occasionally appear on the PT878 display. The
possible Error Messages, along with tips for isolating and correcting
the problem, are listed in Table 3-1 below.
Note: On a graphical display, errors are indicated by a vertical gray
bar at the point of the error. On a numeric display, errors are
indicated by an error code in place of the measurement.
Table 3-1: Error Code Messages
Possible Cause
Displays briefly after clearing
another error message
Broken cable
Flowcell problem
Transducer problem
Electronic failure
Poor flow conditions
Incorrect programming
Bad transducer spacing
Measurement is very different
from the programmed sound
speed
Programming error
Poor flow conditions
Bad transducer spacing
If too high - electronic failure
If too low - flowcell, electrical
problem
Excessive particles or bubbles
in fluid
High second phase percentage
Poor flow conditions
Bad transducer spacing
Under current output
Supply temperature input out of
range
Return temperature input out of
range
Action
None required - measurement is
valid.
Check transducer cable
Check programming and
transducer spacing
Contact the factory if problem
persists.
Check sound speed against
nominal sound speed
Check programming and
transducer spacing
Make sure flow rate is within
±12 m/s (±40 ft/s)
Check programming and
transducer spacing
Check for source of electrical
interference
Check electronics with a test
flowcell
If unit still fails to operate,
contact the factory
Check for flowcell problems
Check transducer spacing as
programmed and as set on pipe
Check that output load is <550
ohm
Check cable and transmitter
Check cable and transmitter
Operation