Modbus holding registers

A.7 Pulse output

Modbus
Data type /
register
Size [bytes]
4:0402
uint8 / 1
4:0077
date / 6
A.7

Pulse output

Table A- 8
Pulse output
Modbus
Data type /
register
Size [bytes]
4:0374
uint8 / 1
4:0375
uint8 / 1
4:0376
float32 / 4
4:0378
uint8 / 1
4:0379
uint8 / 1
50
Parameter
Description
Flow direction totalizer
Calculation principle on flow
2
direction for forward - reverse or
net flow
Totalizer 2 changes
Date and time when totalizer 2
date
has been changed
Parameter
Description
Output A enable
Select 'Yes' to enable output A
Pulse A direction
Pulse output calculation on pure
forward or reverse flow - or net
flow for forward and reverse
pulse flow
Amount per pulse A
Volume per pulse on output A
Pulse width for pulse
Duration of active pulse
A
Output B enable
Select 'Yes' to enable output B
0 = Forward
1 = Reverse
2 = Net
0 = No
1 = Yes
0 = Forward
1 = Reverse
2 = Forward net
3 = Reverse net
0 = 10 ms
1 = 50 ms
2 = 100 ms
3 = 500 ms
4 = 5 ms
5 = 25 ms
0 = No
1 = Yes
Operating Instructions, 04/2016, A5E03409989-AB
Default value
Access level
[unit] / Value
range
Reverse [-]
Read / Write
- [-]
Read Only
Default value
Access level
[unit] /
Value range
Yes [-]
Read / Write
Forward [-]
Read / Write
-
[-]
Read / Write
1)
2)
50 [ms]
Read / Write
Yes [-]
Read / Write