Siemens SITRANS F Operating Instructions Manual page 57

Communication modules sitrans f m mag 8000 modbus rtu
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Modbus
Data type /
register
Size [bytes]
4:0361
uint16 / 2
4:0053
uint16 / 2
MAG 8000 Modbus RTU
Operating Instructions, 04/2016, A5E03409989-AB
Parameter
Description
Alarm configuration
2 bytes describing which er-
list
rors/warning should generate an
alarm or call up on output B
Non optimal measure
Total hours of non optimal
time
measurement conditions
Bit 1: Insulation error
Bit 2: Coil current error
Bit 3: Preamplifier overload
Bit 4: Database checksum
error
Bit 5: Low power warning
Bit 6: Flow overload warning
Bit 7: Pulse A overload warning
Bit 8: Pulse B overload warning
Bit 9: Consumption interval
warning
Bit 10/L: Leakage warning
Bit 11/E: Empty pipe warning
Bit 12/C: Low impedance
warning
Bit 13/d: Flow limit warning
Bit 14/A: Reverse flow warning
Bit 15: Not used
Bit 16: Not used
Modbus holding registers
A.9 Error messages
Default value
Access level
[unit] /
Value range
- [-]
Read Only
0 [h]
Read Only
55