Siemens SITRANS F Operating Instructions Manual page 59

Communication modules sitrans f m mag 8000 modbus rtu
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Modbus
Data type /
register
Size [bytes]
4:0369
uint8 / 1
4:3015
uint32 / 4
4:3031
uint8 / 1
4:0090
uint32 / 4
4:3032
uint8 / 1
4:0370
uint8
4:3043
float32
4:0092
date
4:0823
uint8
: Dependent on order
1)
MAG 8000 Modbus RTU
Operating Instructions, 04/2016, A5E03409989-AB
Parameter
Description
Battery power
Number of installed batteries
(Internal battery pack has 2
batteries and external battery
pack has 4 batteries)
Excitations no.
Numbers of excitations since last
battery reset
Actual battery capacity
Actual remaining battery power
capacity as a percentage of max
capacity
Battery operating time Operating time since installation
of batteries (last battery reset)
Power status
Battery alarm limit
Battery capacity level (%) where
low power alarm will be activated.
Transmitter tempera-
Temperature inside the trans-
ture
mitter (°C)
Battery installation
Latest installation date of batteries
date
Communication mod-
Communication module type -
ule type
for calculating correct power use.
0 = Normal operation
1 = Battery alarm. Actual
battery capacity is below bat-
tery alarm level (% of max
capacity)
2 = Too low power (enters
stand by mode)
3 = As value 1 and 2 together
4 = External power gone
5 = As value 1 and 4 together
6 = As value 2 and 4 together
7 = As value 1 and 2 and 4
together
0 = No module
1 = RS 485
2 = RS 232
3 = RS 232 always connected
Modbus holding registers
A.10 Power control
Default value
Access level
[unit] /
Value range
-
[-]
Read Only
1)
- [-]
Read Only
- [%]
Read Only
0 [h]
Read Only
- [-]
Read Only
10 [-]
Read / Write
- [-]
Read Only
- [-]
Read Only
- [-]
Read Only
57