LeCroy AP034 Instruction Manual page 19

Active differential probe
Table of Contents

Advertisement

Equivalent
Oscilloscope
Scale
Scale Factor
Factor
2 mV/Div
2 mV/Div
5 mV/Div
5 mV/Div
10 mV/Div
10 mV/Div
20 mV/Div
20 mV/Div
50 mV/Div
50 mV/Div
100 mV/Div
100 mV/Div
200 mV/Div
20 mV/Div
500 mV/Div
50 mV/Div
1 V/Div
100 mV/Div
2 V/Div
100 mV/Div
AP034-OM-E Rev D
signal that can be viewed at any equivalent scale factor is the
Input Differential Mode Range plus the offset used. See Adding
Offset on the following page.
When using the AP034 with the ADPPS power supply on
oscilloscopes without ProBus interface, use the following table to
select the desired effective volts per division and determine the
offset available.
Table 2. Recommended AP034 and Oscilloscope settings to obtain desired
AP034
Common
Attenuation
Mode Range
÷1
±16V
÷1
±16 V
÷1
±16 V
÷1
±16 V
÷1
±16 V
÷1
±16 V
÷10
1
±42 V
÷10
1
±42 V
÷10
1
±42 V
÷20
1
±42 V
1
Attenuation obtained using the external ÷10 or ÷20 attenuator.
2
Input differential mode range displayed is limited by ±4 divisions of vertical
scale on oscilloscope.
3
Limited by probe output and ±4 divisions of vertical scale on oscilloscope.
Oscilloscope OFFSET and POSITION must be set to zero.
ISSUED: January 2000
equivalent Volts/Division
Input
Maximum
Differential
Offset
Mode Range
±8 mV
2
±1.6 V
±20 mV
2
±1.6 V
±40 mV
2
±1.6 V
±80 mV
2
±1.6 V
±200 mV
±1.6 V
2
±400 mV
±1.6 V
3
±800 mV
2
±16 V
±2 V
2
±16 V
±4 V
3
±16 V
±8 V
3
±32 V
2SHUDWLRQ
Maximum
Input Noise
Observable
(nV/√Hz),
Signal with
Typical
Offset
±1.608 V
35
±1.620 V
35
±1.640 V
35
±1.680 V
35
±1.800 V
35
±2.0 V
35
±16.8 V
350
±18 V
350
±20 V
350
±40 V
700