DESIGNING TEST FIXTURES FOR THE AP034 PROBE
²
successful completion of the Autobalance cycle. In many
situations, this will be adequate to make routine measurements.
Often it is desirable to connect the probe directly to user-
fabricated test fixtures, such as those used for semiconductor
characterization. To facilitate use with custom test fixtures, the
input receptacles used in the AP034 probe are compatible with
commercially available 0.025 in. (0.635 mm) square pins. The
receptacles do not require a specific rotational orientation for the
square pin. The dimensions listed below can be used as a layout
guide for a test fixture circuit board. The recommended insertion
depth of the pins is 0.100 in. (2.5 mm) to 0.200 in. (5.0 mm).
Top View
Looking toward circuit board
0.105"
0.110"
0.100"
Figure 5. Layout dimensions for test fixtures
# # #
ISSUED: January 2000
$3 $FWLYH 3UREH
0.100"-
0.200"
0.170"
0.025" REF.
Side View
AP034-OM-E Rev D