TEST EQUIPMENT REQUIRED

²
grounding wrist strap and follow standard static
control procedures.
The probe tip housing provides physical rigidity
to the input pins of the probe. When the covers
are removed, observe extra caution to avoid
breaking the probe tip receptacles when mating
the probe to the calibration fixture.
The table on the next page lists the test equipment and
accessories, or their equivalents, that are required for complete
calibration. Specifications given for the test equipment are the
minimum necessary for accurate calibration. All test equipment is
assumed to be correctly calibrated and operating within the
specifications listed. Detailed operating instructions for the test
equipment are not given in this procedure. Refer to the test
equipment manual if more information is needed.
If alternate test equipment is substituted, control settings or
calibration equipment setups may need to be altered. Alternate
models of test equipment may have different connector styles
requiring adapters not included in the equipment list.
ISSUED: January 2000
$3 $FWLYH 3UREH
AP034-OM-E Rev D