Interconnect Cable
Tools

PROCEDURE

AP034-OM-E Rev D
from the output to provide the trigger signal for the test
oscilloscope.
This is for connecting the output of the signal source to the
probe. A BNC cable and a pair of small alligator clips or "lead
grabber" adapter (Pomona #3788) may be used.
0.025 in. (0.635 mm) Square Pins (3 required). The pins from the
header supplied in the probe accessory kit are suitable.
Flat bladed screwdriver, 0.040 in. (1 mm) wide
Adjustment Tool: 0.025 in. (0.635 mm) square head
You can fabricate the Adjustment Tool by flattening the
end of a 0.025 in. (0.635 mm) square pin with a file.
Then, using heat from a soldering iron, insert the pin into
a short length of rigid plastic tubing to serve as a handle.
1. Attach the AP034 to the test oscilloscope. If the test
oscilloscope is not equipped with ProBus, use the ADPPS to
provide power for the AP034.
2. Attach the ÷10 or ÷20 Attenuator Adapter to the AP034 probe
tip.
3. Insert 0.025 in. (0.635 mm) square pins into the +, –, and
input connectors of the Plug-on Attenuator.
4. Attach the interconnect cable to the output of the signal
source.
5. Attach the Trigger Out signal from the signal source to the
External Trigger Input of the test oscilloscope. If the signal
source does not have a separate Trigger Out signal, use a
BNC Tee connector in the output. Run one cable to the
External Trigger Input of the test oscilloscope. Connect the
other to the probe inputs.
ISSUED: January 2000
&DUH DQG 0DLQWHQDQFH
Note
²