Introduction to Programming
Figure 4-7. Logging SCPI Commands within the System Overview Window.
To view commands, select "Read/Clear Remote I/O Traffic Log" and then select
"Generate Report" after the desired configuration has been set. Logged com-
mands are listed in the order received by the instrument. Commands may be
listed individually or as part of a command string.
Note that when changing the 34945EXT module, bank, or channel configura-
tion using the windows shown in Figure 4-2, the SCPI commands associated
with all parameters set by the given window are logged. For example, chang-
ing the channel settling time to 15 ms using the bank 2 channel 1 configuration
window causes the following string of commands to be logged and sent to the
instrument (34945A is in 34980A slot 2):
ROUT:CHAN:LAB "SW 1",(@2121);:ROUT:CHAN:DRIV:PAIR:MODE
ON,(@2121);:ROUT:CHAN:VER:ENAB ON,(@2121);:ROUT:CHAN:VER:POL
NORM,(@2121);:ROUT:CHAN:DRIV:TIME:REC 0.0,(@2121);SETT
0.015,(@2121);:ROUT:CHAN:DRIV:PULS:MODE ON,(@2121);WIDTH
0.015,(@2121);:ROUT:CHAN:DRIV:OPEN:DEF (@2121)