Agilent Technologies 34945A User Manual page 21

Microwave switch/attenuator driver module
Table of Contents

Advertisement

Figure 1-7. Y1159A 16-pin to 16-pin Connector Orientation and Cable Routing.
3. Move the connectors towards the ends of the cable, with some overlap for
strain relief. Note the orientation of the connectors and cable in the figure.
4. Align the metal contacts of the ribbon connector over cable pin 1 and the
other pins, press the ribbon cable connector onto the cable using a vise or
clamp to press the contacts evenly through the cable insulation. Continue to
press the connector until the metal contacts are no longer visible.
5. Repeat Step 4 for the other connector.
6. Fasten the strain relief clamps over the cable and into the ribbon cable
connectors as shown in Figure 1-7. Remove any excess cable after the clamps
are in place.
Agilent 34945A Microwave Switch/Attenuator Driver Module User's Guide
2
1
15
16
About the Instrument
pin 1 of ribbon cable
align connector contacts
over cable lines
"press" until metal contacts
are no longer visible
1
13

Advertisement

Table of Contents
loading

Table of Contents