Figure 1-5. Y1158A 14-pin to 10-pin Connector Orientation and Cable Routing.
3. Move the connectors towards the ends of the cable, with some overlap for
strain relief. Note the orientation of the connectors and cable in the figure.
4. Align the metal contacts of the 14-pin ribbon connector over cable pins 1
through 10. The 14-pin connector will only make contact with pins 1 through
10. Press the ribbon cable connector onto the cable using a vise or clamp to
press the contacts evenly through the cable insulation. Continue to press the
connector until the metal contacts are no longer visible.
5. Repeat Step 4 for the 10-pin to 10-pin connector, noting the location of pin 1
on the cable.
6. Fasten the strain relief clamps over the cable and into the ribbon cable
connectors as shown in Figure 1-5. Remove any excess cable after the clamps
are in place.
Agilent 34945A Microwave Switch/Attenuator Driver Module User's Guide
2
1
only pins 1- 10 are used
14
13
About the Instrument
10-pin connector
pin 1 of ribbon cable
align connector contacts
over cable lines
"press" until metal contacts
are no longer visible
1
11