Table 4
Equipment Required to Test Digital Channel Threshold Accuracy
Equipment
Digital Multimeter
Precision Source
Adapter (qty 2)
BNC Tee
Ω
50
BNC Cable
BNC Test Connector,
8-by-2
Probe Cable
1 Turn on the test equipment and the oscilloscope. Let them warm up for 30
2 Set up the precision source.
3 Use the 8-by-2 test connector and the BNC cable assembly to connect digital
2000/3000 X-Series Oscilloscopes Service Guide
Critical Specifications
0.1 mV resolution, 0.005% accuracy
DC voltage of -5.5 V to 5.5 V, 10 mV resolution
BNC(f) to banana (m)
BNC Tee (m) (f) (f)
BNC - BNC, 48" length
No substitute
minutes before starting the test.
a Set the precision source to provide a DC offset voltage at the Channel 1
output.
Note: Set the Low Force terminal of the Precision Source to its "Floating"
state to prevent offset error caused by ground loop current from the
Precision Source ground to the DUT ground.
b Use the multimeter to monitor the precision source DC output voltage.
channels D0-D7 to one side of the BNC Tee. Then connect the D0-D7 ground
lead to the ground side of the 8-by-2 connector. See
Testing Performance
Recommended Model/Part
Keysight 34401A/34461A
Keysight B2912A/B2962A
Keysight 1251-2277
Keysight 1250-0781 or Pomona 3285
Keysight 8120-1840
User-built (See
page
21)
16-chanel: Keysight N6450-60001 or
N2756-60001
8-channel: Keysight N6459-60001 or
N2755-60001
Figure
2.
2