Information On Electromagnetic Compatibility - Vapormed MIGHTY MEDIC Instructions For Use Manual

Table of Contents

Advertisement

10. THE M IGHTY M EDIC VAPORIZER

10.5. Information on Electromagnetic Compatibility

Guidance and Manufacturer's Declaration – Electromagnetic Emis-
sions
The MIGHTY MEDIC Vaporizer is intended for use in the electromagnetic
environment specified below. The customer or the user of the MIGHTY MED-
IC Vaporizer should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Complies
flicker emissions
IEC 61000-3-3
Electromagnetic environment - guidance
The MIGHTY MEDIC Vaporizer uses
RF energy only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
The MIGHTY MEDIC Vaporizer is
suitable for use in all establishments,
including domestic establishments and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
40
10. THE M IGHTY M EDIC VAPORIZER
Guidance and Manufacturer's Declaration – Electromagnetic Immu-
nity
The MIGHTY MEDIC Vaporizer is intended for use in the electromagnetic
environment specified below. The customer or the user of the MIGHTY MED-
IC Vaporizer should assure that it is used in such an environment.
Immunity
IEC 60601
test
test level
Electrostatic
± 6 kV contact
discharge
discharge
(ESD)
IEC 61000-4-2
± 8 kV air
discharge
Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV line(s) to
IEC 61000-4-5
line(s)
± 2 kV line(s) to
earth
Compliance
Electromagnetic envi-
level
ronment – guidance
± 6 kV contact
Floors should be wood,
discharge
concrete or ceramic
tile. If floors are
± 8 kV air
covered with synthetic
discharge
material, the relative
humidity should be at
least 30%.
± 2 kV for power
Mains power quality
supply lines
should be that of a
typical residential,
not applicable:
commercial or hospital
input/output
environment.
lines are
non-existent
± 1 kV line(s) to
Mains power quality
line(s)
should be that of a
typical residential,
± 2 kV line(s) to
commercial or hospital
earth
environment.
41