8. HYGIENE
8.3. Cleaning of the Cooling
Unit with Mouthpiece
Clean the sink and the surround-
ing area.
Disassemble the Cooling Unit
with Mouthpiece (see Chapter 9.
"Disassembly and Assembly of
the MIGHTY MEDIC Cooling Unit",
page 34).
Remove stubborn residue inside
the Cooling Unit and the Mouth-
piece (if necessary, with ethyl
alcohol).
Then place all pieces of the Cool-
ing Unit with Mouthpiece in warm
(40°C-50°C/104°F-122°F) soapy
water for at least 5 min. (com-
mercial rinsing agent with as few
additives as possible, used ac-
cording to manufacturer's recom-
mendations, avoiding excessive
foam) and ensure that all parts
are completely submerged (and
contain no air bubbles).
Use a soft brush (bottle brush)
or a soft, clean cloth to remove
impurities that is used only for this
purpose, never metal brushes or
steel wool. Clean until there are no
visible impurities on the parts.
Remove the parts from the soapy
water and rinse them for at least
one minute under running water.
Allow the liquid inside to run out
completely and shake any remain-
ing water out of the parts. Avoid
any contact with unclean surfaces,
especially the floor.
Note: If the tap water is very hard, re-
sulting in the formation of a white film
on the surface of the parts, we recom-
mend replacing the Cooling Unit with
Mouthpiece more frequently.
As an alternative to warm water with
wash-up liquid, all parts of the Cool-
ing unit with Mouthpiece (exception:
Cap Lock) can be cleaned with etha-
nol.
Manual cleaning of the disassembled
Cooling Unit
32
8. HYGIENE
Warning! Alcohol is flam-
mable and may easily ig-
nite.
Do not immerse any plastic
components in alcohol for
longer than half an hour. Ex-
tended contact with alcohol may lead
to discoloration or embrittlement of
plastic parts.
8.4. Inspecting/Maintenance of the
Cooling Unit with Mouthpiece
After cleaning, check all component
parts of the Cooling Unit with Mouth-
piece for surface damage, tears, sof-
tening or hardening, soiling, discol-
oration, or excessive limescale and
discard a damaged Cooling Unit or
Mouthpiece.
Then reassemble the Cooling Unit
with Mouthpiece (see Chapter 9.
"Disassembly and Assembly of the
MIGHTY MEDIC Cooling Unit", page
34).
Reusability of the Cooling Unit
with Mouthpiece
The Cooling Unit with Mouthpiece is
intended for multiple use.
The Cooling Unit with Mouthpiece
must be cleaned:
If there is residue (condensate) on
the inside of the Cooling Unit with
Mouthpiece.
At the latest 7 days after the last
use.
Before every use, check the Cool-
ing Unit with Mouthpiece for damage
and wear and tear. A defective Cool-
ing Unit with Mouthpiece must be re-
placed immediately.
Sets of new Cooling Units
with Mouthpiece may be or-
dered from www.vapormed.
com.
Storage
Store Cooling Unit with Mouthpiece in
a dry, dust-free place away from con-
tamination.