Heating Up; Temperature Setting; Reset To Default Settings; Default Settings - Vapormed MIGHTY MEDIC Instructions For Use Manual

Table of Contents

Advertisement

5. COMMISSIONING AND OPER ATION OF THE
M IGHTY M EDIC VAPORIZER
Mains Operation
The MIGHTY MEDIC Vaporizer can
also be operated with the power
adapter, even when the batteries are
completely discharged. Proceed as
described in Chapter 5.4. "Charging
the Battery", page 18 and then
switch on the Vaporizer.
Automatic Switching Off
The automatic switching off is used
to save battery capacity so that the
MIGHTY MEDIC Vaporizer can be
used as long as possible without
charging.
The device switches off automatically
two minutes after the last press of a
button or the last application. The
switching off of the MIGHTY MEDIC
Vaporizer is indicated by a short vibra-
tion.
A short press of the ON/OFF
switch during operation can
reset the automatic switching-
off to two minutes.

5.5. Heating Up

The MIGHTY MEDIC Vaporizer pro-
vides two temperature levels that are
set in the factory: The basic tempera-
ture at 180°C (356°F) and the booster
temperature at 195°C (383°F).
Switch on the device (heating up the
MIGHTY MEDIC Vaporizer with insert-
ed Dosing Capsule and Cooling Unit
mounted is recommended). To do this,
press the orange ON/OFF switch for
at least half a second. The Vaporizer
responds with a short vibration. The
vibration does not take place if there
is a fault.
The display signals that heating is in
progress by showing the set and ac-
tual temperatures. Set the desired
temperature (see Chapter 7. "The Use
of Hemp Flowers (Cannabis Flos)",
page 25) using the temperature
buttons to the left and right of the set
temperature display. Heating up the
heat exchanger can take up to two
minutes, depending on the set tem-
perature and battery charge level.
Once the set temperature is reached,
this is signaled by a short double vi-
bration.
If the basic temperature can no longer
provide adequate evaporation, the
booster temperature can be activated
by means of a double-click on the or-
ange-colored ON/OFF switch. Once
the booster temperature is reached,
this is signaled by a short double vi-
bration and a blinking Set Tempera-
ture display.
20
5. COMMISSIONING AND OPER ATION OF THE
M IGHTY M EDIC VAPORIZER

5.6. Temperature Setting

Set temperature display
(displayed blinking:
booster temperature)
Minus temperature button
The temperature can be set between
40°C and 210°C (104°F and 410°F)
(see Chapter 7. „The Use of Hemp
Flowers
(Cannabis
Flos)",
page
25).
When you press the plus button, the
set value is increased, and when you
press the minus button, the set value
is decreased. When you briefly tap
on the plus or minus button, the set
values are increased or decreased
respectively in 1-degree increments.
If you keep the plus or minus button
pressed, the set values start running
continuously in the corresponding di-
rection.
The booster temperature can be ad-
justed in the same way. However, it
cannot be set lower than the previous
set temperature.
The current actual temperature can be
read from the upper line of the display.
Actual temperature display
Plus temperature button
Battery charging display
Fahrenheit/Celsius
If you simultaneously tap on the plus
button and the minus button, the dis-
play changes from degree Celsius to
degree Fahrenheit and vice versa.

5.7. Reset to Default Settings

It is possible to reset settings. The
reset is made by pressing the ON/
OFF button for ten seconds while
the MIGHTY MEDIC Vaporizer is
switched on. If the ON/OFF button
has been pressed for ten seconds,
the signaling of the reset is made by
vibration. The basic temperature is re-
set to 180°C (356°F) and the booster
temperature is reset to 195°C (383°F).

5.8. Default Settings

Basic temperature:
180°C (356°F)
Booster temperature:
additional 15°C (27°F),
so corresponding to
195°C (383°F)
21

Advertisement

Table of Contents
loading