Vapormed MIGHTY MEDIC Instructions For Use Manual page 23

Table of Contents

Advertisement

10. THE M IGHTY M EDIC VAPORIZER
Immunity
IEC 60601
test
test level
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Compliance
Electromagnetic envi-
level
ronment – guidance
Where P is the
maximum output power
rating of the transmitter
in watts [W] according
to the transmitter
manufacturer and d
is the recommended
separation distance
in meters [m].
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey
, should be
1)
less than the com-
pliance level in each
frequency range
2)
Interference may occur
in the vicinity of
equipment marked with
the following symbol:
44
10. THE M IGHTY M EDIC VAPORIZER
1) Field strengths from fixed transmit-
ters, such as base stations for radio
(cellular/cordless) telephones and
land mobile radios, amateur radio,
AM and FM radio broadcast and TV
broadcast cannot be predicted theo-
retically with accuracy. To assess the
electromagnetic
to fixed RF transmitters, an electro-
magnetic site survey should be con-
sidered. If the measured field strength
in the location in which the MIGHTY
MEDIC Vaporizer is used exceeds
Recommended Separation Distances between portable and mobile RF
Communications Equipment and the MIGHTY MEDIC Vaporizer
The MIGHTY MEDIC Vaporizer is intended for use in an electromagnetic en-
vironment in which radiated RF disturbances are controlled. The customer or
the user of the MIGHTY MEDIC Vaporizer can help prevent electromagnetic
.
interference by maintaining a minimum distance between portable and mo-
bile RF communications equipment (transmitters) and the MIGHTY MEDIC
Vaporizer as recommended below, according to the maximum output power
of the communications equipment.
Rated maximum
output power of
transmitter
[W]
0,01
0,1
1
10
100
the applicable RF compliance level
above, the MIGHTY MEDIC Vaporizer
should be observed to verify normal
operation. If abnormal performance is
observed, additional measures may
be necessary, such as re-orienting
or relocating the MIGHTY MEDIC Va-
environment
due
porizer.
2) Over the frequency range 150 kHz
to 80 MHz, field strengths should be
less than 3 V/m.
Separation distance according to frequency of
transmitter [m]
150 kHz – 80 MHz
80 MHz – 800 MHz
d = 1,2 √P
d = 1,2 √P
0,12
0,12
0,38
0,38
1,2
1,2
3,8
3,8
12
12
45
800 MHz – 2,5 GHz
d = 2,3 √P
0,23
0,73
2,3
7,3
23

Advertisement