9. DISASSEMBLY AND ASSEMBLY OF THE
M IGHTY M EDIC COOLING UNIT
Turn the Cooling Unit with a 90° ro-
tation counterclockwise in order to
remove it from the MIGHTY MEDIC
Vaporizer.
Pull out the Mouthpiece from the Cool-
ing Unit Cap, rotate slightly back and
forth while doing this, and pull off the
Mouthpiece Seal Ring.
Pulling the Mouthpiece from the Cooling Unit
Cap
Open the Cap Lock by pulling it rear-
wards.
Opening the Cap Lock
Remove the Cap Lock from the cap by
pulling up at one side.
If the Cooling Unit must be im-
mersed into alcohol (ethanol)
the Cap Lock only has to be
removed from the Cooling Unit. Ex-
tended contact with alcohol can lead
to whitening of the Cap Lock Label or
disintegration of the glue.
Removal of the Cap Lock
Separate the Cooling Unit Cap from
the Cooling Unit Base by pulling off
the cap upwards.
Pull the Cooling Unit Cap from the Cooling
Unit Base
34
9. DISASSEMBLY AND ASSEMBLY OF THE
M IGHTY M EDIC COOLING UNIT
Remove the large and small Base
Seal Rings. Lever out the large Seal
Ring from the recess in the notch us-
ing the Filling Chamber Tool.
Removal of the large Base Seal Ring
Push the upper Screen downwards
out of the Filling Chamber Cap from
above by using the Filling Chamber
Tool supplied.
Pressing out the upper Screen
Reassemble in reverse order.
When assembling, make sure
that Seal Rings are properly
seated. The Mouthpiece Seal
Ring, especially, must sit completely
in the cap and be no longer visible.
The upper Screen must be inserted
carefully into the groove provided.
Disregarding this may allow plant
particles to enter the Cooling Unit, so
blocking this or be inhaled.
When using the Dosing Capsules dur-
ing operation, the Lower Screen will
normally not get polluted. If the screen
does have to be exchanged one day,
proceed as follows:
Remove the lower Filling Chamber
Screen out of the Filling Chamber by
using a pair of tweezers or a pair of
pointed pliers. Insert a new Screen
while taking care that the edges of the
Screen seat into the groove at the bot-
tom of the Filling Chamber.
35