INSTRUCTIONS FOR USE
Read carefully before using the device
and save these instuctions.

Advertisement

Table of Contents
loading

Summary of Contents for Vapormed MIGHTY MEDIC

  • Page 1 INSTRUCTIONS FOR USE Read carefully before using the device and save these instuctions.
  • Page 2: Table Of Contents

    3. Intended Use .............. 13 4. Important Remarks ............. 14 Operating Conditions, Storage and Transport Conditions ...... 16 5. Commissioning and Operation of the MIGHTY MEDIC Vaporizer ..........17 5.1. Unpacking ............17 5.2. Power Connection ..........17 5.3. Application and Operation ........18 5.4.
  • Page 3: Product Overview, Scope Of Delivery, Service

    8.4. Inspecting/Maintenance of the Cooling Unit with Mouthpiece ......33 9. Disassembly and Assembly of the MIGHTY MEDIC Cooling Unit ........34 10. The MIGHTY MEDIC Vaporizer ........36 10.1. Cleaning of the MIGHTY MEDIC Vaporizer and the Power Adapter ......36 Instructions 10.2.
  • Page 4 Actual Temperature Display Please check whether all components have been included. If not, please inform our Service Center. Operating life: The MIGHTY MEDIC Vaporizer has an average expected operating life of: Minus Temperature Plus Temperature – approx. 1,000 operating hours...
  • Page 5: Explanation Of Symbols, Safety Recommendations

    2. EXPLANATION OF SYMBOLS, SAFETY RECOMMENDATIONS 2. EXPLANATION OF SYMBOLS, SAFETY RECOMMENDATIONS Please read the following important to ensure safety during The MIGHTY MEDIC Vaporizer is protected by an angle of 15° installation, use and maintenance safety recommendations against dropping water (IP 22). The device should nevertheless be thoroughly and carefully of the MIGHTY MEDIC Vaporizer.
  • Page 6 (such as UL, IMQ, VED, make sure that you always keep +S etc.) may be used, provided the the MIGHTY MEDIC Vaporizer and UL Recognized Component Mark: This symbol indicates that the specified power exceeds the power the accessories out of the reach of...
  • Page 7: Intended Use

    15° against dropping water (IP paired or transported when the power 22). The device should nevertheless adapter is plugged in. The MIGHTY MEDIC Vaporizer is in- be protected against moisture and The Vaporizer and power adapter tended for vaporizing and then inhal- rain.
  • Page 8: Important Remarks

    The MIGHTY MEDIC Vaporizer is not responding laws. cist or manufacturer of medical hemp intended for use in children and youth flowers (Cannabis flos).
  • Page 9: Operating Conditions

    +35°C +60°C erence. er type label corresponds to the local power supply voltage. Remove the MIGHTY MEDIC Vapor- izer and the accessories from the box. All parts of the electrical system must +5°C -20°C Immediately after unpacking, check...
  • Page 10: Application And Operation

    Note: When the power adapter is off the MIGHTY MEDIC Vaporizer and cating the charging status. If the bat- Connecting the MIGHTY MEDIC Vaporizer plugged into the mains, the dis-...
  • Page 11: Heating Up

    The minutes, depending on the set tem- When you press the plus button, the switching off of the MIGHTY MEDIC perature and battery charge level. set value is increased, and when you 5.7. Reset to Default Settings...
  • Page 12: Dosing Capsules With Magazine

    Dosing Cap- tation counterclockwise in order to pre-filled in Dosing Capsules by the sules is filled and insert them into the remove it from the MIGHTY MEDIC nursing staff. magazine. The magazine can hold up Vaporizer.
  • Page 13: The Use Of Hemp Flowers (Cannabis Flos)

    Filling Aid is available as ers for authenticity, purity, and patho- produced by the MIGHTY MEDIC Va- an accessory, see Chapter 11. “Spare gens according to the directives of the porizer.
  • Page 14: Preparation

    (vapor) recognizable when exhaling. The Dosing Capsule’s content is tation counterclockwise in order to then depleted. remove it from the MIGHTY MEDIC Using the MIGHTY MEDIC Vaporizer Vaporizer remove the Dosing Capsule from the Filling Chamber, open it and If the vapor decreases when exhaling 7.1.
  • Page 15: Hygiene

    Cooling Unit with Mouthpiece and new and in a low-germ, but not fully- Condensation or residual moisture infection from a contaminated the MIGHTY MEDIC Vaporizer do sterile state. may encourage bacterial growth. Lip Piece, a contaminated Mouth- not come in contact with the following Its purpose is to avoid a health risk –...
  • Page 16: Cooling Unit With Mouthpiece

    Afterwards, it essary (see Chapter 8.3. „Cleaning Product Overview of the MIGHTY MEDIC Cooling Unit must be disposed of in the household of the Cooling Unit with Mouthpiece“, The Cooling Unit consists of: waste and a new Lip Piece must be page 32).
  • Page 17: Cleaning Of The Cooling Unit With Mouthpiece

    “Disassembly and Assembly of tended contact with alcohol may lead If there is residue (condensate) on the MIGHTY MEDIC Cooling Unit“, As an alternative to warm water with to discoloration or embrittlement of the inside of the Cooling Unit with page 34).
  • Page 18: Disassembly And Assembly Of The Mighty Medic Cooling Unit

    (ethanol) Seal Rings. Lever out the large Seal When assembling, make sure remove it from the MIGHTY MEDIC the Cap Lock only has to be Ring from the recess in the notch us- that Seal Rings are properly Vaporizer.
  • Page 19: The Mighty Medic Vaporizer

    Possible Cause / Remedy under running water. of children or unqualified persons. MIGHTY MEDIC cannot be Please ensure that the MIGHTY MEDIC’s battery is In a physician’s office or clinic, the switched on. charged or the Vaporizer is connected via the power...
  • Page 20: Technical Specifications

    Patents and designs: www.vapormed.com/patents Batteries are specially assembled consumable items that can only be replaced by our Service Center when the MIGHTY MEDIC Vaporizer is sent in to them. Any attempt by the customer to replace the battery by him/herself is potentially dangerous and leads to loss of any claims under the warranty and exempts Storz &...
  • Page 21: Information On Electromagnetic Compatibility

    Guidance and Manufacturer’s Declaration – Electromagnetic Immu- nity Guidance and Manufacturer‘s Declaration – Electromagnetic Emis- The MIGHTY MEDIC Vaporizer is intended for use in the electromagnetic sions environment specified below. The customer or the user of the MIGHTY MED- The MIGHTY MEDIC Vaporizer is intended for use in the electromagnetic IC Vaporizer should assure that it is used in such an environment.
  • Page 22 – guidance nity The MIGHTY MEDIC Vaporizer is intended for use in the electromagnetic Voltage dips, <5% Uτ (>95% <5% Uτ (>95% Mains power quality environment specified below. The customer or the user of the MIGHTY MED- short dip in Uτ) for...
  • Page 23 Communications Equipment and the MIGHTY MEDIC Vaporizer electromagnetic site survey , should be The MIGHTY MEDIC Vaporizer is intended for use in an electromagnetic en- less than the com- vironment in which radiated RF disturbances are controlled. The customer or pliance level in each...
  • Page 24: Spare Parts And Accessories

    (6 pcs. Upper Screen with coarse mesh width) propagation is affected by absorption and reflection from REF 04 04 MM MIGHTY MEDIC Mouthpiece Set (4 pcs. Mouthpiece) structures, objects and people. MIGHTY MEDIC Seal Ring Set (3 pcs. Mouthpiece Seal Ring, REF 06 01 MM 3 pcs.
  • Page 25: Declaration Of Ce Conformity

    12. DECL AR ATION OF CE CONFORMIT Y 13. WARR ANT Y, LIABILIT Y, REPAIR SERVICE The MIGHTY MEDIC Vaporizer and Warranty covered to warranty law through re- power adapter fulfill the requirements pair or by delivering a new product to Storz &...
  • Page 26: Proof Of The Annual Safety Checks

    13. WARR ANT Y, LIABILIT Y, REPAIR SERVICE 14. PROOF OF THE ANNUAL SAFET Y CHECKS ( SC ) Repair Service Date SC Finding Comments Signature Name / Title In addition to our warranty obligations, we offer competent repair service to our customers.
  • Page 27 S TO R Z & B I C K E L G M B H & C O. K G In Grubenäcker 5-9 · 78532 Tuttlingen / Germany Phone +49-74 61- 96 9707- 0 · Fax +49 -74 61-96 97 07- 7 e-mail: info@vapormed.com...