10. HOT AIR GENER ATOR
If the preceding tips do not
yield the desired results, or
for any other defects or prob-
lems not mentioned here, please pull
the power plug and contact our Ser-
vice Center immediately.
Do not open the Hot Air Gen-
erator! Without special tools
and specialized knowledge,
any attempt to open the device will
cause damage to it. Such an attempt
will invalidate the warranty claim.
Do not try to remedy the defect your-
self. Contact directly our Service
Center.

10.7. Periodic Safety Checks

The following safety checks should be
performed at least every 24 months
by a qualified person who has ade-
quate training, knowledge, and practi-
cal experience to perform these tests.
Inspect the device and accesso-
ries for mechanical and functional
damage.
Inspect the safety labels for leg-
ibility.
Verify that the device functions
properly as described in the
Instructions for Use.
The safety check is to be recorded in a
device log and data of control results
are to be documented.
If the device is not functioning prop-
erly or fails any of the above tests, the
device has to be repaired.

10.8. Technical Specifications

VOLCANO MEDIC
Vaporization System
Voltage:
220-240 V / 50-60 Hz (alternatively
available with 110-120 V / 50-60 Hz)
The specification is located on the
bottom side of the VOLCANO MEDIC
Hot Air Generator.
Power consumption: 120 VA
Air pump output: 10 VA
Air temperature: adjustable between
approx. 40 - 210°C (104 - 410°F)
Air flow: approx. 12 l/min
Valve Balloon Size:
approx. 12.5 liters
Size of the Hot Air Generator: 20.0 x
18.0 cm (7.9 x 7.1 inch)
Weight of the Hot Air Generator with
power cord, Filling Chamber, and
Valve Balloon with Mouthpiece: ap-
prox. 1.8 kg (3.9 lbs)
Safety category II
Please unplug to disconnect from the
mains current.
Internal fuses F1 and F2: T 1.25 A
(applies to 110-120 V and 220-240 V
devices)
These fuses may be replaced only by
our Service Center.
48
10. HOT AIR GENER ATOR
Subject to technical changes.
Patents and designs:
www.vapormed.com/patents
10.9. Information on Electromagnetic Compatibility
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The VOLCANO MEDIC Vaporization System is intended for use in the electro-
magnetic environment specified below. The customer or the user of the VOL-
CANO MEDIC Vaporization System should assure that it is used in such an en-
vironment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Complies
flicker emissions
IEC 61000-3-3
Manufacturer
Storz & Bickel GmbH & Co. KG
In Grubenäcker 5-9,
78532 Tuttlingen,
Germany
Electromagnetic environment - guidance
The VOLCANO MEDIC Vaporization
System uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The VOLCANO MEDIC Vaporization
System is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.