10. THE M IGHTY M EDIC VAPORIZER
Immunity
IEC 60601
test
test level
Voltage dips,
<5% Uτ (>95%
short
dip in Uτ) for
interruptions
½ cycle
and voltage
variations on
40% Uτ (60%
power supply
dip in Uτ) for 5
input lines
cycles
IEC 61000-4-11
70% Uτ (30%
dip in Uτ) for
25 cycles
<5% Uτ
(>95% dip in
Uτ) for 5 s
Note: Uτ is the a.c. mains voltage prior to application of the test level.
Compliance
Electromagnetic envi-
level
ronment – guidance
<5% Uτ (>95%
Mains power quality
dip in Uτ) for
should be that of a
½ cycle
typical residential,
commercial or hospital
40% Uτ (60%
environment. The
dip in Uτ) for 5
battery powered
MIGHTY MEDIC
cycles
Vaporizer is able to offer
70% Uτ (30%
continuing functionality
dip in Uτ) for
even if the energy
25 cycles
supply is interrupted.
<5% Uτ
(>95% dip in
Uτ) for 5 s
42
10. THE M IGHTY M EDIC VAPORIZER
Guidance and Manufacturer's Declaration – Electromagnetic Immu-
nity
The MIGHTY MEDIC Vaporizer is intended for use in the electromagnetic
environment specified below. The customer or the user of the MIGHTY MED-
IC Vaporizer should assure that it is used in such an environment.
Immunity
IEC 60601
test
test level
Conducted RF
3 V
rms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5
GHz
Compliance
Electromagnetic envi-
level
ronment – guidance
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the MIGHTY
MEDIC Vaporizer,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance:
d = 1,2 √P
3 V
rms
d = 1,2 √P
3 V/m
für 80 MHz to 800 MHz
d = 2,3 √P
for 800 MHz to 2,5 GH
43