6. Troubleshooting
6.2 Error code
Error
No display
None
No running
Unknow Rotor ID!
Not open lid
None
Not close the lid
To close the door
perfectly.
During
** Imbalance
Acceleration,
Error! **
the device
vibrates and
switch off.
Failure of power
** Power
Failure! **
Etc.
** Motor
Overheat! **
** Over Speed! ** - Run above 105% of the set spped
* If none of these solutions work, please send an email to techsupport@ihanil.com.
18
www.ihanil.com
Display
- No power connection.
- Failure of power supply
- No rotor
Errors in drive or rotor
-
recognition function.
- Foreign substances in
the rotor or sensor.
-Fauiure of power supply
- No working lid pin
- Lid is not closed tightly.
- Not load symmetrically
- Not tighten the rotor with shaft
- The device is placed uneven place
- Failure of power supply
- Motor over heat
Cause
Check the power cable
-
Check the main fuse
-
in laboratory and device
- Load a rotor.
- Turn off the main power and
turn on again.
- Remove foreign matter from
the rotor or sensor.
- Stop the running and open
the lid by emergency key.
- Press the Door button to open
the lid and then close it again.
- Turn off the main power and
turn on again.
- Check the rotor and samples
- Tighten the rotor properly.
- Place the device on flat place
- Check a outlet
- Stop the centrifugation and
lower the temperature
- Stop and run again
Solution
.