Hach DR/700 Procedures Manual page 665

Colorimeter
Table of Contents

Advertisement

OXYGEN, DISSOLVED, SHR,
SUMMARY OF METHOD
The High Range Dissolved Oxygen AccuVac Ampul contains reagent
vacuum sealed in a 12-mL ampul. When the AccuVac Ampul is broken
open in a sample containing dissolved oxygen, it forms a yellow color
which turns purple. The purple color development is proportional to the
concentration of dissolved oxygen.
REQUIRED REAGENTS
Description
High Range Dissolved Oxygen
AccuVac Ampuls, with
2 reusable ampul caps . . . . 1 ampul . . . . 25/pkg . . . . . . . . 25150-25
REQUIRED APPARATUS
Adapter, AccuVac Vial . . . . . . 1 . . . . . . . . . each . . . . . . . . . . 46025-00
Beaker, 50 mL . . . . . . . . . . . . . 1 . . . . . . . . . each . . . . . . . . . . . . 500-41
Caps, Ampul, blue . . . . . . . . . . 1 . . . . . . . . . 6/pkg . . . . . . . . . . 1731-06
DR/700 Filter Module
Number 69.01. . . . . . . . . . . 1 . . . . . . . . . each . . . . . . . . . . 46269-00
OPTIONAL APPARATUS
Dissolved oxygen may also be determined by titrimetric methods.
Request Publication 8042 for additional information.
AccuVac Snapper Kit . . . . . . . . . . . . . . . . . . each . . . . . . . . . . 24052-00
AccuVac Dissolved Oxygen Sampler . . . . . . each . . . . . . . . . . 24051-00
BOD Bottle and Stopper, 300 mL . . . . . . . . . each . . . . . . . . . . . . 621-00
Cap for 10- and 25-mL sample cells . . . . . . . 12/pkg . . . . . . . . 24018-12
Sample Cell, 10-mL with screw cap . . . . . . . 6/pkg . . . . . . . . . 24276-06
Sample Cell, 25-mL with screw cap . . . . . . . 6/pkg . . . . . . . . . 24019-06
For Technical Assistance, Prices and Ordering
In the U.S.A. - Call 800-227-4224 toll-free for more information.
Outside the U.S.A. - Contact the Hach office or the distributor serving you.
continued
Quantity
Per Test
Unit
69-5
Cat. No.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents