OXYGEN, DISSOLVED, HR,
SUMMARY OF METHOD
The High Range Dissolved Oxygen AccuVac Ampul contains reagent
that is vacuum sealed in a 12-mL ampul. When the AccuVac ampul is
broken open in a sample containing dissolved oxygen, it forms a yellow
color which turns purple. The purple color development is proportional
to the concentration of dissolved oxygen.
REQUIRED REAGENTS
Description
High Range Dissolved Oxygen
AccuVac Ampuls, with 2
reusable ampul caps . . . . . . 1 ampul . . . . 25/pkg . . . . . . . . 25150-25
REQUIRED APPARATUS
Beaker, 50 mL . . . . . . . . . . . . . 1 . . . . . . . . . each . . . . . . . . . . . . 500-41
Caps, ampul, blue . . . . . . . . . . varies . . . . . . 6/pkg . . . . . . . . . . 1731-06
DR/700 Filter Module
Number 52.01. . . . . . . . . . . 1 . . . . . . . . . each . . . . . . . . . . 46252-00
OPTIONAL APPARATUS
AccuVac Snapper Kit . . . . . . . . . . . . . . . . . . each . . . . . . . . . . 24052-00
Adapter, AccuVac Vial . . . . . . . . . . . . . . . . . each . . . . . . . . . . 43784-00
BOD bottle and stopper, 300 mL . . . . . . . . . each . . . . . . . . . . . . 621-00
Cap for 10- and 25-mL sample cells . . . . . . . 12/pkg . . . . . . . . 24018-12
Dissolved Oxygen Sampler,
AccuVac Model . . . . . . . . . . . . . . . . . . . . each . . . . . . . . . . 24051-00
Sample Cell, 10-mL with screw cap . . . . . . . 6/pkg . . . . . . . . . 24276-06
Sample Cell, 25-mL with screw cap . . . . . . . 6/pkg . . . . . . . . . 24019-06
Dissolved oxygen may also be determined by titrimetric methods.
Request Publication 1171 for additional information.
For Technical Assistance Prices and Ordering
In the U.S.A. - Call 800-227-4224 toll-free for more information.
Outside the U.S.A. - Contact the Hach office or distributor serving you.