PALLADIUM,
13.
Fill a 10-mL
cell to the 10-mL line
with the blank. Cap.
16.
Fill a 10-mL
sample cell with 10
mL of the prepared
sample.
continued
14.
Place the blank
in the cell holder.
Note: In bright light, it
may be necessary to
close the cell
compartment cover.
17.
Place the
prepared sample in
the cell holder.
Note: In bright light, it
may be necessary to
close the cell
compartment cover.
42-53
15.
Press:
ZERO
The display will
count down to 0.
Then the display will
show 0 mg/L and the
zero prompt will turn
off.
18.
Press:
READ
The display will
count down to 0.
Then the display will
show the results in
mg/L palladium (Pd).