Hach DR/700 Procedures Manual page 365

Colorimeter
Table of Contents

Advertisement

CADMIUM,
15 seconds
7.
Add 20 mL of
50% Sodium
Hydroxide Solution
and the a 0.1-g scoop
of potassium cyanide
to the funnel. Shake
vigorously for 15
seconds. Remove the
stopper and let stand
for one minute.
Note: Spilled reagent
will affect test accuracy
and is hazardous to skin
and other materials.
continued
60 seconds
8.
Add 30 mL of
DithiVer Solution to
the 500-mL
separatory funnel.
Stopper, invert, and
open the stopcock
slowly to vent. Close
the stopcock and
shake the funnel once
or twice; vent again.
Close the stopcock
and shake the funnel
vigorously for 60
seconds.
52-39
1 minute
9.
Let the funnel
stand undisturbed for
about 1 minute.
Note: The bottom layer
(chloroform) will be
pink if cadmium is
present.