Loading Slides - Beckman Coulter UniCel DxH User Manual

Analysis system
Table of Contents

Advertisement

UniCel DxH Slidemaker Stainer
Coulter Cellular Analysis System
The software supply icon turns yellow when the slide level is low.
The software supply icon will turn red when the slides are completely out. The system
will go offline until more slides are loaded. Go back online to continue processing the
slides.
Risk of injury and biohazardous contamination. Handle slides with care to avoid skin puncture.
Clean up any broken glass or blood spill as quickly as possible. Dispose of all contaminated
disposable cleaning materials and broken slides in accordance with your local regulations and
acceptable laboratory practices.
NOTE: There are 2 sensors on the slide chute. One of the sensors, checks the slide inventory below ½
(approx 75 slides). When this happens the Supplies Icon will be in Yellow. The other sensor will stop the
Slidemaker Stainer. This will occur when there are 12 slides left in the chute. The Supplies Icon will be in
red.
Step
Open the slide ejector door.
1
Remove slides from the box handling them by the painted end.
2
PN: B11956AA (March 2012)
Professional Development

Loading Slides

Action
Page 37 of 46