Product installation

Environmental operating
requirements
ECO8000 Series Automatic Changeover Unit User Manual
Check that the location of your installation has the proper operating environment
as listed in the following table. (See Table 2.)
CAUTION.
Damage to the instrument can occur if this instrument is powered on at
temperatures outside the specified temperature range.

Table 2: ECO8000 Series environmental requirements

Parameter
Temperature
Operating
Nonoperating
Relative
Operating
Humidity
Nonoperating
Altitude
Operating
Nonoperating
Leave space for cooling by ensuring standard side clearance for rack mounting or
2 inches (5.1 cm) of side clearance for benchtop use. Also, ensure sufficient rear
clearance (approximately 2 inches) so that cables are not damaged by sharp bends.
For complete specifications for the instrument, refer to the ECO8000 Series
Specifications and Performance Verification Technical Reference. You can locate
this document on the Tektronix Web site (www.tektronix.com\downloads).
Description
0 °C to +50 °C
–20 °C to +60 °C
20% to 80% (No condensation); Maximum wet-bulb
temperature 29.0 °C
5% to 90% (No condensation); Maximum wet-bulb
temperature 40.0 °C
To 3,000 m (9,842 feet)
Maximum operating temperature decreases 1 °C for
each 300 m above 1.5 km.
To 15,000 m (49,212 feet)
Getting Started