ECO8000 Series Automatic Changeover Unit User Manual
4. When the message Select upgrade network mode: is displayed, press the
left (◄) or right (►) arrow button to select from the following modes and
then press the ENTER button.
DHCP – Use this selection if the instrument is connected to a network
with DHCP support. An IP address will be assigned to the instrument
from your network server.
Local IP (192.168.1.1) – Use this selection if the instrument is connected
to a network without DHCP support or is connected to a standalone
computer. In this case, the factory default, static IP address of 192.168.1.1
will be used for the upgrade.
NOTE.
When you select either the Local IP or the Saved static IP upgrade
mode, you cannot edit the displayed IP address. The IP address cannot be
edited when the instrument is in recovery mode.
Quit and continue booting – Use this selection to quit the upgrade
process and leave the instrument in its current condition.
5. If you did not select Quit and continue booting in the previous step, the top
line of the instrument display will show the message Waiting for network
connection and the second line will list an IP address. Make a note of the
displayed IP address.
6. Enter the IP address you noted in the previous step into the upgrade transfer
application window on your PC to restart the upgrade process.
CAUTION.
To prevent upgrade failure, leave the instrument power on during the
upgrade process. The upgrade process may take up to 15 minutes to complete.
7. When the upgrade is complete, the instrument should reboot. If it does not,
cycle the power.
If you continue to experience an upgrade problem, contact Tektronix customer
support or your local Tektronix representative for assistance.
Operating Basics
47