Understanding Error and Other Messages
Message
What It Means
A used test strip
was inserted.
The blood or control
solution sample was
applied before the
symbol
appeared.
The temperature
during the test was
above or below the
operating range.
What To Do
Repeat the test with a
new test strip.
Repeat the test with a
new test strip and
wait until the
symbol appears
before applying the
blood or control
solution sample.
Move to an area
where the
temperature is within
the operating range
(10-40°C) and repeat
the test after the
meter and test strips
have reached a
temperature within
the operating range.
www.caresens.co.nz 41