Caution: Storage and Handling

• Do not expose the meter to direct sunlight or heat for
extended periods of time.
• Do not let dirt, dust, blood, or water enter into the meter's test
strip port.
• Do not drop the meter or subject it to strong shocks.
• Do not try to fix or alter the meter in any way.
• Keep the meter in a cool and airy place.
• Keep the meter away from strong electromagnetic field
sources such as cell phones and microwave ovens.
• The CareSens N Voice Meter should be used only with
CareSens N Single Test strips.
• Store all meter components in the carrying case to prevent
loss.
Note:
You can get additional information or technical
assistance by calling our Customer Service Center: 1-800-429-
5001 (9 am - 9 pm EST, Monday - Saturday).
46 www.i-sens.com

Understanding Error and Other Messages

Message
What is said
"Error 1. Used
A used test strip
test strip."
was inserted.
"Error 2. The
The blood or
sample has
control solution
been applied
sample was
before the
applied before
blood symbol
the
appeared."
appeared.
N/A
The thermometer
stabilized yet.
"Error 4. The
The blood
sample was
sample has
not enough or
abnormally high
too thick."
viscosity or
insufficient
volume.
What it means
What to do
Remove the strip
and insert a new
test strip.
Repeat the test
with a new test
strip and wait until
the
symbol
symbol
appears before
applying the
blood or control
solution sample.
Retry the test
of meter is not
after about 30
minutes for the
thermometer to
be stabilized.
Repeat the test
using a new test
strip.
www.i-sens.com 47