The performance of CareSens N Blood Glucose Monitoring
System has been evaluated in laboratory and in clinical tests.
Accuracy: The accuracy of the CareSens N BGM System (Model
GM505NA, GM505NB, GM505NC) was assessed by comparing
blood glucose results obtained by patients with those
obtained using a YSI Model 2300 Glucose Analyzer, a laboratory
instrument. The following results were obtained by diabetic
patients at clinic centers.
Slope
Y-intercept
Correlation coefficient (r)
Number of sample
Range tested
Accuracy results for glucose concentration < 4.2 mmol/L
(Within ± 0.28 mmol/L )
(Within ± 0.56 mmol/L ) (Within ± 0.83 mmol/L )
13/17 (76%)
16/17 (94%)
Accuracy results for glucose concentration ≧ 4.2mmol/L
Within ± 5%
Within ± 10%
45/93 (48%)
88/93 (96%)
44 www.i-sens.com
0.961
0.194 mmol/L
0.995
110
1.7~26.9 mmol/L
17/17 (100%)
Within ± 15%
Within ± 20%
93/93 (100%)
93/93 (100%)
Precision: The precision studies were performed in a laboratory
using CareSens N BGM Systems.
Within Run Precision
Blood average
2.1 mmol/L
Blood average
4.8 mmol/L
Blood average
6.9 mmol/L
Blood average
10.5 mmol/L
Blood average
18.6 mmol/L
Total Precision
Control average
2.4 mmol/L
Control average
6.3 mmol/L
Control average
21.2 mmol/L
This study shows that there could be variation of up to 4.2%
SD = 0.1 mmol/L
SD = 0.2 mmol/L
CV = 4.1%
CV = 2.6%
CV = 2.8%
SD = 0.1 mmol/L
CV = 3.3%
CV = 4.2%
www.caresens.co.nz 45