Comparing the Control Solution Test Results

The test result of each control solution should be within the
range printed on the label of the test strip vial. Repeat the
control solution test if the test result falls outside of this range.
Out of range results may occur due to the following factors:
Situations
• W hen the control solution bottle was
not shaken well,
• W hen the meter, test strip, or the control
solution were exposed to high or low
temperatures,
• W hen the first drop of the control
solution was not discarded or the tip of
the bottle was not wiped clean,
• W hen the meter is not functioning
properly.
• W hen the control solution is past the
expiration date printed on the bottle,
• W hen the control solution is past its
discard date (the date the bottle was
opened plus three (3) months),
• W hen the control solution is
contaminated.
If results continue to fall outside the range printed on the
test strip vial, the CareSens N Test Strip and Meter may not be
working properly. Do no use your system and contact i-SENS
sales representative.
22 www.i-sens.com

Using the Lancing Device

You will need a lancing device in order to collect a blood
sample.
You may use the lancing device contained in the CareSens
N Blood Glucose Monitoring System or any other medically
approved lancing device.
Actions
Repeat the
control solution
test by referring
to the "Notes"
on page 19.
• T he lancing device may not be used by more than one
• U se a soft cloth or tissue to wipe the lancing device.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.
Lancet Holder
Adjustable Tip
Lancet
individual. Ensure the lancing device is not shared among
different users.
If necessary, a small amount of alcohol on a soft cloth or
tissue may be used.
Caution:
To avoid infection when drawing a sample, use a
lancet only one time, and:
• D o not use a lancet that has been used by others.
• Always use a new sterile lancet.
• Keep the lancing device clean.
Note:
Repeated puncturing at the same sample site may
cause pain or skin calluses (thick hard skin). Choose a
different site each time you test.
Ejection System
Release Button
Sliding Barrel
www.caresens.co.nz 23