Understanding Error and Other Messages
Message
What It Means
A used test strip
was inserted.
The blood or control
solution sample was
applied before the
icon
appeared.
The temperature
during the test was
above or below the
operating range.
40 www.i-sens.com
What To Do
Repeat the test with a
new test strip.
Repeat the test with a
new test strip and
wait until the
icon appears before
applying the blood
sample.
Move to an area
where the
temperature is within
the operating range
(10 ~ 40 °C/ 50 ~ 104
°F) and repeat the test
after 30 minutes.
Message
What It Means
The blood sample
has abnormally high
viscosity or
insufficient volume.
A non-CareSens N
test strip was used.
Test strip was not
inserted properly.
There is a problem
with the meter.
Note:
If the error messages persist, contact your authorized
i-SENS sales representative.
What To Do
Repeat the test after
inserting a new test
strip.
Repeat the test with a
CareSens N test strip.
Insert a test strip with
the contacting bars
facing upwards and
push in gently until
the meter beeps.
Do not use the meter.
Contact your
authorized i-SENS
sales representative.
www.i-sens.com 41