Accuracy Table - BRAND Transferpette S Operating Manual

Hide thumbs Also See for Transferpette S:
Table of Contents

Advertisement

S , adjustable volume
Transferpette
®
Volume range
µl
0.1 - 1
0.1 - 2.5
0.5 - 10
2 - 20
5 - 50
10 - 100
20 - 200
100 - 1000
500 - 5000
1000 - 10000
* A = Accuracy, CV = Coefficient of Variation
!
Final test values related to the nominal capacity (maximum volume)
indicated on the instrument, and the indicated volume steps obtained when
instrument and distilled water are equilibrated at ambient temperature
(20 °C/68 °F) and with smooth operation. According to DIN EN ISO 8655.
Volume step
A*
µl
≤ ± %
1
0.5
0.1
20
2.5
1.25
0.25
12
10
5
1
20
10
2
50
25
1.2
5
4
100
0.6
50
0.8
10
3
200
0.6
100
0.8
20
3
1000
0.6
500
0.8
100
3
5000
0.6
2500
0.8
500
3
10000
0.6
5000
0.8
1000
3
20 °C
Ex
CV*
Increment
≤ %
µl
2
1.2
0.001
4
2.4
12
1.4
0.7
0.002
2.5
1.5
6
1
0.5
0.01
1.6
1
7
4
0.8
0.4
0.02
1.2
0.7
5
2
0.8
0.3
0.05
0.5
2
0.2
0.1
0.4
1
0.2
0.2
0.3
0.6
0.2
1
0.3
0.6
0.2
5
0.3
0.6
0.2
10
0.3
0.6

Accuracy Table

Recommended
type of tip, µl
0.1 - 20
0.5 - 20
0.5 - 20
2 - 200
2 - 200
2 - 200
2 - 200
50 - 1000
500 - 5000
1000 - 10000
37