Pipetting

5 ml and 10 ml instruments should only be used with the PE filter
installed (see page 40).
Pipette tips are disposables items!
32
1.
Fitting the tip
Use the correct tip according to the volume range
or the color code. Ensure that the tip is securely
seated.
2.
Volume setting
a) Push the volume-change protection upward to
b) Select the desired volume by rotating the volume-
setting wheel. Avoid twisting and abrupt rotating
motions during this adjustment.
c) Push the volume-change protection down to
protection tightens but does not lock volume-
setting wheel.
3.
Aspirate sample
a) Press pipetting button to the first stop.
b) Hold the pipette vertically and immerse
the tip into the liquid.
Volume
range
0.1 µl - 1 µl
> 1 µl - 100 µl
> 100 µl - 1000 µl
> 1000 µl
Immersion
Waiting
depth in mm
time in s
1 - 2
2 - 3
2 - 4
3 - 6
1
1
1
3