Limitations of Use

Purpose
The pipette is an air-displacement system for pipetting aqueous solutions
with medium density and low to medium viscosity.
Limitations of Use
The instrument is intended for the pipetting of liquids within the following
limitations:
– use temperature from +15 °C to +40 °C (59 °F to 104 °F). Consult
the manufacturer for use in temperatures outside of this range.
– Vapor pressure up to 500 mbar
– Viscosity: 260 mPa s (260 cps)
Operating Limitations
Viscous and highly adhesive liquids may impair volumetric accuracy.
Volumetric accuracy may also be impaired when pipetting liquids that
differ from ambient temperature by more than ± 1 °C /± 1.8 °F.
Operating Exclusions
The user has to ensure the compatibility of the instrument with the
intended application.
This instrument cannot be used:
for liquids attacking polyphenyl sulfide (PPS) (50 µl pipet, adjustable
volume)