Problem
System didn't
accept recent
calibration.
No readings,
Alarm, or Alerts
display until
error is fixed.
System didn't
accept recent
calibration.
No readings,
Alarm, or Alerts
display until
error is fixed.
What you do
Wait 15 minutes.
Enter 1 calibration.
If error screen still
appears, enter 1 more
meter value.
Wait 15 minutes.
If no readings display
on the smart device or
receiver, the sensor needs
to be replaced.
Tap error for more
information.
To report error, contact
Tech Support (see
beginning of section).
95
Verify you didn't enter
a meter value outside
the range of 40 and
400 mg/dL.
Wait until meter value is
40-400 mg/dL.