distinguishes from the value shown while the transducer
was coupled, step 4 is to be repeated.
6) The
and the
key are to be used to adjust the
material thickness up and down until it matches the
material thickness of the sample piece.
7) The
key has to be pressed to confirm. The hint will
change to "NO2" indicating to be ready to measure the
second calibration point.
8) Steps 3 to 7 are to be repeated. The hint will change
back to "DPC".
The instrument is now ready to perform measurements
within its range.

4.7 Scan mode

While the instrument excels in making single point
measurements, it is sometimes necessary to examine a
larger region, searching for the thinnest point. This
instrument includes a feature, called SCAN- Mode, which
allows to do just that.
During normal operation, it performs and displays four
measurements every second which is adequate for single
measurements. In SCAN- Mode, however, the instrument
performs ten measurements every second and displays
the readings while scanning. While the transducer is in
contact with the material to be measured, it is always
keeping track to finding the lowest measurements. The
transducer may be "scrubbed" across the surface, any
brief interruptions of the signal will be ignored. If it looses
contact with the surface for more than two seconds, the
instrument will display the smallest measurement it found.
On the 【 Test Set 】 - > 【 Work mode 】 menu item the
key has to be pressed to toggle between single point mode
and scan mode.
4.8 Limit set
With the Limit set feature the user is able to set an audible
and visual parameter while taking measurements.
If a measurement is beyond the limit range, set by the
user, the beeper will sound, if enabled. With this, the
speed and efficiency of the inspection process is improved
by elimination of constant viewing of the actual reading
displayed. In the section below it is described how to set
up this feature:
1) On the 【 Test Set 】 - > 【 Tolerance Limit 】 menu item the
key has to be pressed to activate the limit set screen.
2) The
key, the
and
change the bottom limit and the upper limit value to the
desired values.
3) The
key has to be pressed to confirm the change
and to return to the previous screen, or the
be pressed to cancel the change.
TU_US-BA-e-1110
Sauter GmbH
Tieringerstr. 11-15
D-72336 Balingen
E-Mail: info@sauter.eu
Instruction Manual
TU_US
key are to be used to
key has to
Tel: +49-[0]7433- 9976-174
Fax: +49-[0]7433-9976-285
Internet: www. sauter.eu
If the measurement range is exceeded, the user will be
reminded to re-set. If the bottom limit is larger than the
upper limit, the values will be exchanged automatically.
4.9 Changing resolution
The instrument has got a selectable display resolution,
which is 0.1 and 0.01mm.
On the 【 Test Set 】 - > 【 Resolution 】 menu item the
has to be pressed to switch between "high" and "low".
4.10 Unit scale
On the 【 Test Set 】 - > 【 Unit 】 menu item the
to be pressed to switch back and forth between imperial
and metric units.
4.11 Memory management
4.11.1 Storing a reading
There are 20 files (F00-F19) which can be used to store
the measurement values inside the instrument.
At most 99 records can be stored in each file.
The following steps outline how to do this:
1) The
key has to be pressed to activate the
【 File Name 】 item on the main measurement screen.
2) The
and
key are to be used to select the desired
file to save the data.
3) After a new measurement reading appears, the
has to be pressed to save the measurement value to the
current file. If the 【 Auto Save 】 function is activated, the
measurement value will be automatically saved to the
current file after a new measurement operation.
4.11.2 Viewing a stored reading
The
key has to be pressed multiple times until the
【 Record No 】 area on the measurement screen is
highlighted. The
and
the record number. The stored values, according to the
record number from the memory, will be automatically
read. They are displayed on the "Thickness reading" area
of the screen simultaneously.
If the displayed record data shall be deleted from the
memory, the
key has to be pressed.
The other way to view the stored record is by menu
operations. On the 【 Memory Manager 】 - > 【 View Mem
Data 】 menu the
key has to be pressed to activate the
memory viewing screen.
key
key has
key
key are to be used to change
7