5 Operation

5.1 Setting the working mode

Users and inspectors often deal with coated materials such as pipes and tanks on site.
Typically, inspectors must remove the paint or coating prior to measurement or take
into account a certain number of defects caused by paint or coating thickness or speed.
The error can be eliminated with this meter by using a special echo-echo mode
designed for measurements in such cases. The gauge is equipped with this user-
friendly feature, so there is no need to remove the paint or coating.
To switch between P-E mode and E-E mode, press [Test Settings] in the Test Settings
dialog box.
5.2 Probe selection
Make sure that you have set the correct probe model in the device. Otherwise, the
measurement may be incorrect. In the Probe Model dialog box, use the
to scroll to the currently used probe model.
Then press the
and exit the dialog box.

5.3 Function "Probe Zero"

The key
is used to zero the instrument in the same way as a mechanical micrometer
is zeroed. If the gauge is not set to zero correctly, some error may occur in all
measurements taken with the gauge. When the Meter is zeroed, this fixed error value
is measured and automatically corrected in all subsequent measurements. Zeroing the
instrument can be done in the following ways:
1. Connect the converter to the device. Make sure that the connectors are fully
engaged. Make sure that the wear surface of the converter is clean and free of
any foreign matter.
2. Press the button
3. Apply a single drop of the coupling agent to the surface of the metal plate of the
probe.
TO-BA-e-2020
or F3 key to confirm the selection. Press
to activate the Probe Zero mode (see figure below).
and keys
to cancel the operation