display can be decreased/ increased with the
key until the desired value is matched.

5.5 Save and exit

The
key has to be pressed to confirm the modification
and to return back to the previous screen.
5.6 Cancel and exit
The
key has to be pressed to cancel the change and
return back to the previous screen.
6. Servicing
If there should appear some abnormal phenomena to the
instrument, please do not dismantle or adjust any fixed
assembly parts on your own. Instead of this, the present
warranty card has to be filled out and the instrument has to
be sent to us. The warranty service can be carried on.
7. Transport and Storage
The instrument has to be kept away from vibration, strong
magnetic fields, corrosive medium, dumpiness or dust.
Storage in ordinary temperature.
Appendix A Sound velocities
Material
In/us
Aluminum
0.250
0.233
Steel, common
0.226
Steel, stainless
0.173
Brass
0.186
Copper
0.233
Iron
0.173-0.229
Cast Iron
0.094
Lead
0.105
Nylon
0.142
Silver
0.128
Gold
0.164
Zinc
0.236
Titanium
0.117
Tin
0.100
Epoxy resin
0.157
Ice
0.222
Nickel
0.106
Plexiglass
0.092
Polystyrene
0.230
Porcelain
0.094
PVC
0.222
Quartz glass
0.091
Rubber, vulcanized
0.056
Teflon
TU_US-BA-e-1110
Sauter GmbH
Tieringerstr. 11-15
D-72336 Balingen
E-Mail: info@sauter.eu
Instruction Manual
and
Sound Velocity
m/s
6340-6400
5920
5740
4399
4720
5930
4400 - 5820
2400
2680
3607
3251
4170
5990
2960
2540
3988
5639
2692
2337
5842
2388
5639
2311
1422
TU_US
Water
Appendix B Application Notes
Measuring pipe and tubing
When a piece of pipe is measured to determine the
thickness of the pipe wall, the orientation of the transducer
is of importance. If the diameter of the pipe is larger than
approximately
4
performed with the transducer orientated in the way that
the gap in the surface of the sensor is perpendicular (at
right angle) to the long axis of the pipe.
For smaller pipe diameters, two measurements should be
performed, one with the surface gap of the sensor
perpendicular, another with the gap parallelto the long axis
of the pipe. The smaller one of the displayed values should
be taken as the thickness of that point.
Measuring hot surfaces
The sound velocity through a substance is dependent on
its temperature. As materials heat up, the velocity of sound
through them decreases. In most applications with surface
temperatures of less than 100°C, no special procedures
must be observed. At temperatures above that point, the
change in sound velocity of the material being measured
starts
having
a
measurement. At such elevated temperatures it is
recommended to first performing a calibration on a sample
piece of known thickness, which is at or near the
temperature of the material being measured. This will allow
the instrument to correctly calculate the sound velocity
through the hot material.
When performing measurements on hot surfaces, it may
also be necessary to use a specially constructed high-
temperature transducer. These transducers are built of
materials which can withstand high temperatures.
It is also recommended that the sensor has to be left in
contact with the surface for a short time in order to aquire a
stable measurement. While the transducer is in contact
with the hot surface, it will be heated up and with termal
expansion and other effects, the accuracy of measurement
may adversely be affected.
Measuring laminated materials
Laminated materials are unique because of their density
(and therefore sound velocity) may considerably vary from
one piece to another. Some laminated materials may even
exhibit noticeable changes in sound velocity across a
single surface. The only way to a reliable measurement is
to perform a calibration on a sample piece of known
thickness. Ideally, this sample material should be a part of
the same piece being measured, or at least from the same
lamination batch. The effects of variation of sound velocity
Tel: +49-[0]7433- 9976-174
Fax: +49-[0]7433-9976-285
Internet: www. sauter.eu
0.058
1473
inches,
measurement
should
noticeable
effect
upon
be
ultrasonic
10